Introduction

The following guide serves as a checklist for the possible causes and solutions with respect to some of the most commonly encountered problems from the flow cytometry (FACS) experiments. We at Boster Bio are committed to helping our customers “get better results”. While the troubleshooting guide below covers a multitude of problems encountered while performing flow cytometry experiments, we do not expect it to be the exclusive solution to any problems during your specific experiment. We do hope that you will find the information beneficial to you and useful as a reference guide in troubleshooting any FACS problems you may encounter. If you ever need more assistance with your flow cytometry experiments, please contact the Boster Support Team by email at [email protected]

Troubleshooting Guides

Download troubleshootingnhandbooks for IHC, Western blot and ELISA for FREE.

troubleshooting_box_image Troubleshooting guides
  • Weak Fluorescence Intensity or No Fluorescent Signal

    S.No. Possible Cause Solution
    1 The antibodies are degraded or expired
    • Ensure that antibodies are stored as per the instructions of manufacturer.
    • Keep track of antibody stocks; make sure products are not expired.
    • If titrating antibodies and storing aliquots of the same, add sodium azide in the storage buffer at 0.09%.
    2 The fluorescence of the fluorochrome has faded
    • Be sure to store the conjugated antibodies away from light exposure. Fresh antibodies will be needed.
    3 The antibody concentration is too low for detection
    • Titrate the antibodies before use to find the optimal amount to use for your specific experiment.
    • Use negative (unstained) and positive controls.
    4 Expression of target antigen is too low
    • Check literature for antigen expression in different cell types and use a suitable positive control.
    • Whenever possible, use freshly isolated cells as opposed to frozen samples.
    • Optimize cell culture/stimulation protocols in case the antigen (e.g. cytokines) expression depends on prior in vitro treatment.
    5 Antigen-antibody binding is sub-optimal
    • Check the species specificity of the antibody.
    • Optimize the antibody incubation time and temperature.
    • Consider using biotinylated primary antibodies and an additional biotin-steptavidin step to amplify the signal.
    6 The intracellular antigen is not accessible
    • Optimize cell permeabilization protocols.
    7 The intracellular antigen is getting secreted
    • Target must be membrane bound or cytoplasmic to be detected. Try using a Golgi blocker such as Brefeldin A.
    8 The surface antigen is getting internalized
    • Perform all protocol steps at 4°C and use ice cold reagents. We recommend you permeablize on ice.
    9 The fluorescence on stained cells has bleached
    • Acquire the cells immediately after staining.
    • Add a fixative (like PFA) to the samples if storing for an extended duration; alcohol fixatives are best avoided.
    10 A low expressing antigen has been paired with a dim fluorochrome
    • Always pair the weak antigens with bright fluorochromes such as PE or APC.
    11 The primary and secondary antibodies are not compatible
    • Use a secondary antibody that was raised against the species in which the primary antibody was raised.
    12 The laser and PMT settings are not compatible with fluorochrome or PMT voltage is too low for the fluorescent specific channel
    • Ensure that proper instrument settings are loaded prior to acquisition.
    • Use suitable positive and negative controls to optimize settings for every fluorochrome.
    13 The fluorescent signal is over compensated
    • Use MFI alignment instead of visual comparison to compensate.
  • Saturated or Excess Fluorescent Signal

    S.No. Possible Cause Solution
    1 The antibody concentration is too high
    • Titrate the antibodies before use to find the optimal amount to use for your specific experiment.
    • Use appropriate positive and negative controls.
    2 Unbound antibodies are trapped in the cells in the case of intracellular staining
    • Wash the cells adequately after each antibody incubation step and include Tween or Triton X in wash buffers.
    3 A high expressing antigen is paired with bright fluorochrome
    • Always pair strong antigens with dimmer fluorochromes such as FITC or Pacific Blue.
    4 The PMT voltage is too high for the fluorescent specific channel
    • Ensure proper instrument settings are loaded prior to acquisition.
    • Use suitable positive and negative controls to optimize settings for every fluorochrome.
    5 The fluorescent signal is under-compensated
    • Use MFI alignment instead of visual comparison to compensate.
    6 Inadequate blocking
    • Add 1% to 3% blocking agent with antibody as well as a blocking step (dilute antibodies in blocking solution), and increase the blocking time.
  • High Background or Non-specific Staining

    S.No. Possible Cause Solution
    1 Excess, unbound antibodies are present in the sample
    • Wash cells adequately after every antibody incubation step
    2 Non-specific cells are targeted
    • Include an isotype control to subtract any Fc binding signal.
    • Block the Fc receptors on cells with Fc blockers, BSA or FBS prior to antibody incubation.
    • Include additional washing steps.
    • Include a secondary antibody-conjugate control to subtract any non-specific signal from the conjugate.
    • Select a secondary antibody that does not cross react.
    3 High auto-fluorescence
    • Always include an unstained control to subtract the auto-fluorescence signal.
    • For cells with naturally high auto-fluorescence (e.g. neutrophils), use fluorochromes that emit in the red channel where auto-fluorescence is minimal (e.g. APC).
    • If the above solution is not possible, use very bright fluorochromes for these cell types to amplify the signal above the auto-fluorescence level.
    • Avoid storing the cells in a fixative solution for long durations; analyse cells soon after staining if possible.
    4 Presence of dead cells
    • Always sieve the cells once before acquiring and sorting to remove any dead cell debris.
    • Include viability dyes like PI or 7-AAD to gate out any dead cells.
    • Use freshly isolated cells as opposed to frozen cells whenever possible.
  • High Background Scatter or Abnormal Scatter Profile of Cells

    S.No. Possible Cause Solution
    1 The cells are lysed or damaged
    • Optimize sample preparation to avoid cell lysis.
    • If possible, do not vortex or centrifuge cells at high speeds.
    • Use fresh buffers.
    • Avoid storing the stained cells for long durations.
    2 Bacterial contamination
    • Store stained cells and antibodies properly to avoid bacterial growth. Bacteria will exhibit low levels of auto fluorescence.
    • Practice proper sterile cell culture techniques to prevent bacterial contamination.
    3 Incorrect instrument settings for scatter
    • Ensure proper instrument settings are loaded prior to acquisition.
    • Use fresh, healthy cells to correctly set the FSC and SSC settings for that cell type.
    4 Presence of dead cells
    • Always sieve the cells once before acquiring and sorting to remove any dead cell debris.
    • Use freshly isolated cells as opposed to frozen cells whenever possible.
    5 Presence of un-lysed RBCs (red blood cells)
    • Ensure RBC cells lysis is complete – check using a microscope.
    • Use fresh RBC lysis buffer.
    • Wash as many times as needed to remove RBC debris.
    • If using PBMCs after a ficoll gradient, optimize the procedure to minimize RBC contamination in the lymphocyte interphase.
  • Abnormal Event Rate

    S.No. Possible Cause Solution
    1 Event rate is low due to low cell number
    • Keep the minimum cell count at 1X106/ml.
    • Ensure the cells are mixed well with gentle pipetting.
    2 Event rate is low due to sample clumping
    • Always sieve the cells once prior to acquiring and sorting to remove debris.
    • Ensure the cells are mixed well with gentle pipetting before staining, and again before running your samples.
    3 Incorrect instrument settings
    • Adjust the threshold parameter as needed.
    4 No events due to a clogged sample injection tube
    • Unclog flow cytometer injection tube as per instrument manufacturer's instructions (typically run 10% bleach for 5-10 min, followed by dH2O for 5-10 min).
    5 Event rate is too high due to concentrated sample
    • Dilute the cell count to 1X106/ml.
    6 Event rate is too high due to air in flow cells and/or sheath filter
    • Refer to the instrument manual for the appropriate steps.
  • Loss of Epitope

    S.No. Possible Cause Solution
    1 Excessive paraformaldehyde
    • Paraformaldehye can release methanol in its breakdown, which may affect the staining. Make sure to use only 1% paraformaldehyde.
    2 Sample was not kept on ice
    • Keep antibodies at 4 °C to prevent loss of activity. This also prevents active phosphatases and proteases from altering the epitope of interest.
    3 Sample fixed too long
    • Optimize fixation protocol. Fixing the samples for too long may cause damage to cells. Most cells only need to be fixed for less than 15 minutes.

Popular Antibodies

Here are the 300 most popular antibodies.

acetylcholinesterase antibody ADAM10 antibody ADAMTS13 antibody adiponectin antibody
AGO2 antibody AHR antibody ALK antibody alkaline phosphatase antibody
AMH antibody ANG antibody annexin a1 antibody APOE antibody
AQP1 antibody aquaporin 4 antibody ARC antibody ATF4 antibody
ATF6 antibody ATM antibody ATRX antibody BAX antibody
bcl xl antibody BCR antibody beclin 1 antibody BRCA1 antibody
CAD antibody calreticulin antibody caspase 1 antibody caspase 8 antibody
catalase antibody cathepsin b antibody Ccl2 antibody CCR4 antibody
CCR5 antibody cd11b antibody cd11c antibody Cd14 antibody
cd16 antibody CD163 antibody CD19 antibody cd2 antibody
cd20 antibody CD200 antibody CD24 antibody CD27 antibody
cd30 antibody CD33 antibody CD34 antibody CD36 antibody
Cd40 antibody CD44 antibody cd45 antibody CD47 antibody
CD63 antibody CD68 antibody cd8 antibody CD80 antibody
CD81 antibody CD86 antibody Cd9 antibody CDK2 antibody
CDK4 antibody ceruloplasmin antibody CHAT antibody claudin 5 antibody
collagen i antibody collagen ii antibody creb antibody Crp antibody
csf antibody CSF1R antibody Ctla4 antibody CXCL10 antibody
CXCR3 antibody CXCR4 antibody cyclin antibody cytokeratin 5 antibody
desmoglein 3 antibody DKK1 antibody DLL4 antibody DNMT1 antibody
doublecortin antibody EEA1 antibody EGFR antibody enos antibody
EPCAM antibody ERBB2 antibody erythropoietin antibody EZH2 antibody
fak antibody FAP antibody FAS antibody FGF21 antibody
Fgf23 antibody FGFR1 antibody FLT3 antibody FOS antibody
FOXO1 antibody FOXP3 antibody GAL antibody GAPDH antibody
GATA3 antibody GDF15 antibody glut4 antibody gp100 antibody
GPI antibody GPX4 antibody grp78 antibody HGF antibody
HIF1A antibody HLA-A antibody hsp90 antibody huntingtin antibody
iba1 antibody ICAM1 antibody Icos antibody IDS antibody
IFNAR1 antibody Igf1r antibody Il10 antibody IL13 antibody
IL15 antibody Il17a antibody IL18 antibody IL1B antibody
IL2 antibody IL33 antibody IL5 antibody IL6R antibody
IL8 antibody inos antibody IRF3 antibody islet 1 antibody
JAK2 antibody jnk antibody KEAP1 antibody KLF4 antibody
L1CAM antibody lactoferrin antibody lamin a antibody LAMP1 antibody
LAT antibody leptin antibody Lif antibody LOX antibody
lysozyme antibody MAG antibody MAX antibody MBP antibody
MDM2 antibody MERTK antibody mesothelin antibody MICA antibody
MIF antibody MLH1 antibody MMP2 antibody MMP9 antibody
MOG antibody MTOR antibody MUC2 antibody myeloperoxidase antibody
NANOG antibody nephrin antibody nestin antibody NGF antibody
NLRP3 antibody NOTCH1 antibody NOX4 antibody NPY antibody
NRP1 antibody occludin antibody osteocalcin antibody osteopontin antibody
p300 antibody p63 antibody parkin antibody parp antibody
PAX6 antibody PAX8 antibody PCSK9 antibody pd l1 antibody
PDGFRA antibody PDGFRB antibody perilipin antibody perk antibody
Pf4 antibody pgp9.5 antibody PML antibody ppar gamma antibody
PROX1 antibody PTEN antibody rab7 antibody RAC1 antibody
RAD51 antibody rea antibody RET antibody rip3 antibody
RIPK1 antibody RUNX2 antibody SHH antibody SIRT1 antibody
SMAD2 antibody smad3 antibody SNAP25 antibody SOD1 antibody
SOD2 antibody somatostatin antibody SOX10 antibody SOX2 antibody
SOX9 antibody SP1 antibody SPR antibody SRC antibody
STAT3 antibody STAT6 antibody survivin antibody SYK antibody
TAZ antibody TBK1 antibody tdt antibody TERT antibody
thrombin antibody Thy1 antibody tim 3 antibody TLR2 antibody
TLR3 antibody TLR4 antibody Tnf antibody topoisomerase i antibody
transferrin antibody trkb antibody TSG101 antibody TSHR antibody
TSLP antibody ubiquitin antibody VDAC1 antibody vegf antibody
VWF antibody XBP1 antibody ZEB2 antibody